mOuNTING
Install well into customer-supplied 1/2” FNPTsaddle (or Thredolet®).
Screw sensor into the well via the 1/8” NPSM brass tting.
WIrING
Terminate with butt splice or soldering. Wire nuts are not recommended.
Use full 8’ (2.44m) lead length provided to avoid moisture migration up
the wire.
Flow Flow
Angle must not allow end of well
Install the wells as shown to obtain
optimum well strength.
DImENSIONS
1/2" NPT
5.13
(13.03)
3.25
(8.26)
4.87
(12.38)
0.25
(0.63)
1.37
(3.49)
(XL)
1/8" NPSM
Sensor Sensor in Well
Greenfield
Fitting
in
For technical support, call 877-373-2700.
SENSAPHONE
®
REMOTE MONITORING SOLUTIONS
901 Tryens Road
Aston, PA 19014
www.sensaphone.com
mOuNTING
Install well into customer-supplied 1/2” FNPTsaddle (or Thredolet®).
Screw sensor into the well via the 1/8” NPSM brass tting.
WIrING
Terminate with butt splice or soldering. Wire nuts are not recommended.
Use full 8’ (2.44m) lead length provided to avoid moisture migration up
the wire.
Flow Flow
Angle must not allow end of well
Install the wells as shown to obtain
optimum well strength.
DImENSIONS
1/2" NPT
5.13
(13.03)
3.25
(8.26)
4.87
(12.38)
0.25
(0.63)
1.37
(3.49)
(XL)
1/8" NPSM
Sensor Sensor in Well
Greenfield
Fitting
in
For technical support, call 877-373-2700.
SENSAPHONE
®
REMOTE MONITORING SOLUTIONS
901 Tryens Road
Aston, PA 19014
www.sensaphone.com